Savill P. Continuous ECG vs loop recording for AF detection post stroke. Practitioner April 2019;263(1825):8

Continuous ECG vs loop recording for AF detection post stroke

24 Apr 2019

AUTHORS

Cardiovascular disease

Dr Peter Savill BSc (Hons) MBBS PGDipCard, Former GP, Community Cardiology Specialist, Southampton and Mid Hampshire

Article

Automatic external loop recording (ELR) is not suitable as a single monitoring device for AF screening following stroke, a prospective cohort study from Denmark has concluded.

Consecutive patients aged 60 years and over who had had an acute ischaemic stroke or transient ischaemic attack (TIA) with no prior AF were recruited to the study between June 2013 and December 2017. A total of 1,412 patients, mean age 72.8 years, were monitored for 48 hours by simultaneous continuous ECG recording and ELR.

The primary outcome was the detection of AF lasting longer than 30 seconds. The gold standard was taken as cardiologist verified AF in continuous ECG monitoring. The authors also hypothesised ELR to detect 15% fewer patients with AF than continuous ECG recording.

In continuous ECG monitoring 38 (2.7%) patients had AF. ELR automatically categorised 219/1,412 patients (15.5%) with AF, including 32/38 (85%) patients with AF on continuous ECG monitoring. After cardiologist review of ELR recordings, AF was diagnosed in 57/219 patients, of which 32 (56%) had AF on continuous ECG monitoring.

For cardiologist adjudicated AF detection by ELR, sensitivity was 84% (95% CI: 69-94%), specificity was 98% (95% CI: 97-99%) and positive predictive value was 56% (95% CI: 42-69%).

Screening for AF in patients who have had an ischaemic stroke or TIA is important to ensure the timely initiation of oral anticoagulation as prophylaxis against recurrent stroke.

A 12-lead ECG has some use in this situation but if this shows sinus rhythm then more prolonged ECG recording will be required. The optimal method of ECG monitoring is often disputed.

The authors conclude that automatic AF detection with ELR results in an AF diagnosis in more than five patients without AF for each patient with AF as verified on continuous ECG monitoring. However, with cardiologist verification of AF detection by ELR, sensitivity improved to 84% and specificity 98%.

‘Automatic ELR as investigated in this study may be considered to rule out AF but is not suitable as a single monitoring device for AF screening in patients early after stroke,’ they state.

However, ELR is probably of more use in monitoring over more extended periods of time such as seven days or more to increase the yield of AF detection but this was not the intention of this trial.

 

Dr Peter Savill

Sejr MH, May O, Damgaard D et al. External continuous ECG versus loop recording for atrial fibrillation detection in patients who had a stroke. Heart Published Online First: 21 March 2019. doi: 10.1136/heartjnl-2018-314186